In this video, I am going to show how to fix the RF input of a Siglent Spectrum Analyzer. Model number is SSA 3021X.
Full article can be found [ Ссылка ]
Corrections: Front panel is sensitive to low frequency (sub 10 kHz) stress, the voltage level is not 'high voltage' as I wrongly put in the video. I meant 'higher than rated' voltage. But the reality is even exposing this front panel to 0dBm low frequency noise for a long time could cause damage.
Chips need to be replaced:
PIN Diode D31, D33 ------- BAR63-02V(Infineon)
Voltage regulator diode D32, D34 ------- BZX84J-C30(NXP)
GaAs SPDT Switch U1 --------- MASWSS0181TR(M/A-COM)
If it is a Rigol DSA815 unit, chipset is HMC221B
And special thanks to Mr Michael Mayerhofer of Tekbox [ Ссылка ] for providing the repair details.
Happy fixing!