For helium gas measurement, Helium ( He) thermal mass flowmeter must do zero calibrated. The method is below:
1. To install the He flowmeter on the pipeline. When the pipeline is full of helium gas and does not flow, the gas at static condition ;
2. Press the ENT to enter the setting, enter the password 1000, press ENT to confirm, and press the upward arrow to cycle to the ”Zero voltage “ and “ current voltage”, , press ENT twice to make the “Zero Voltage” consistent with “ Current Voltage” , press ESC to exit